TBA-C series, Toshiba's latest automated clinical chemistry analyzer (TBA ™ -c16000, TBA ™ -C8000, TBA ™ -C4000). Compatibility developed and immune instrument, TBA-C series data of the same accuracy can be obtained by a simple operation.
Simple start of analysis at any time, 24 hours a day
Minimized sample carryover, enables new analytical sequences
Low final reaction volume of 80uL
Sample clot detection function for prevention of sample throughput reduction
˙ Minimized sample Carry - over 0.1ppm or less
Sample carryover is reduced to 0.1 ppm or less by the enhanced sample probe wash operation. It is no longer necessary to divide a sample into small portions for use in an immunoassay system.
˙ From Biochmical assays to immunoassays
Biochemical assays and immunoassays have conventionally been performed in seperate dedicated systems. By minimizing the carryover between samples, TBA-c8000 makes it possible to run biochemical assays and immunoassays as part of a continuous assay sequence using a single sample tube. TBA-c8000 thus increase assay throughput to perviously unattainable levels, with more intelligent, efficient, and speedy analysis.
˙ Set and Go - Automatic start of analysis
No rack jams. No result buffer wait time.
Analysis can be started automatically by simply setting a tube containing a sample on the intelligent sampler.
˙ High-preision photometry : Enables measurement using a microvolume
Toshiba'a extensive experience and advenced technologies have been integrated to reduce reaction volume and improve data percision. Photometry is possible for a minimum reaction volume of 80uL for undiluted samples. This feature improves the cost/ performance ratio through reduction of the reagent requirements, and it enables a wide reange of application.
˙ Flex Rate Function: Decrease the retest ratio
For high-concentration samples that require retesting after dilution, the Flex Rate function calculates the concentration based on the absorbance date in the early stages of the read timing and reports the result. This minimizes the retest ratio abd consequently the reporting time and opration cost.
˙ Hard glass reaction cuvettes: Facilitate effective washing
TBA-c8000 employs molded hard-glass reaction cuvettes that can be washed effectively.
˙ Sample Clot Detection function: Contributes to a short reporting time
The sample clot detection unit monitors the preesure during sample aspiration abd dispensing abd detects any abnormalities in the pressure waveform. If cloggings in the sample probe due to fibrin and other obstacles are detected, an error is output and the sample probe ia washed automatically. Sampling of the next sample is stared when the pressure returns to normal levels.
˙ Compact integrated electrolyte measurement system with excellent maintainability
Toshiba's original integrated multi-ion sensor can simultaneously measure three electrolytes (Na, K, Cl) in serum or urine and is designed for easy maintenance. CV < 1.0 %
Max. No. assays | 206 |
Throughput | 1200 Test / Hr |
Methods | End - Point, Rate, and ISE |
Sample carryover | 0.1 ppm or less |
Sample container | sample tubes, sample cups |
Isothermal condition | Water Bath |
Photometry | Direct photometry of reaction cuvettes, monochromatic or bichromatic measurement |
Data processing function |
Autodilution, Auto-calibration, Autorerun, serum Index Multipoint calibration curve, clibartion expiration date, clibration curve log reaction time course (33 points), reagnets/ detergent remaining volumes, Smart wash lamp intensity monitor, Flexrate....... |
Online interface |
RS - 232C |
Dimensions |
2066 ( W ) * 1625 ( D ) * 2017 ( H ) mm |
Mass |
675 Kg |
Power supply |
AC200V ± 10 % |
Power consumption |
3 KVA Maxima |