TBA-FR series is inherited Toshiba technology, the pursuit of high reliability further automatic biochemical analyzer equipment. TBA - 120 FR, Accute (TBA - 40 FR), TBA - 25 FR and TBA ™ - 2000 FR, can meet all the needs of large and small laboratory
˙ Analytical throughput
The accute can perform simultaneous analysis of up to 35 tests at a speed of up to 400 tests / hour using photometric technoloques.
˙ The minimum reaction liquid volume for photometry is 120uL
Photometric analysis can be performed with a minimum reaction liquid volume is 120uL, reducing reagent consumption. The reaction volume can be set in the range from 120uL to 360 uL, to support od a wide range of application.
˙ Each probe contains an integrated level sensor
The sample and reagent probes each contain an integrated level sensor. They detect the sample or reagnet surface, minimizing contact and reducing the interference between different samples and between different reagents.
˙ The durable hard-glass cuvettes can be wash effectively
The Accute uses cast hard-glass cuvettes that are extremely durable. To prepare for the next measurement, the reaction cuvettes are automatically washed with deionized water and detergent during operation.
˙ Automatic sample dilution
The sample is automatically diluted for special chemistry according to the setting.
˙ The stirrer throughly mixes the reaction liquid
Mixing and washing are completed within the measurement cycle time. The simple design of the stirrer ensures easy maintenance.
Max. No. assays | 35 ( ISE up to 38 ) |
Throughput | 400 Test / Hr ( ISE up to 800 Test / Hr ) |
Methods | End - Point, Rate, and ISE |
Sample containers | sample tubes, sample cups |
Isothermal condition | Water Bath |
Photometry | Direct photometry of reaction cuvettes, monochromatic or bichromatic measurement |
Data processing function |
Multipoint calibration curve, clibartion expiration date, clibration curve log,automatic calibration, reaction time course (33 points), reagnets/ detergent remaining volumes, reagent bottle changover, skipping of insufficient reagnet, Smart wash, autodilution, autorerun, serum Index, automation startup/ shutdown, lamp intensity monitor...... |
Online interface | RS - 232 C |
Dimensions | 1030 (W) * 793.6 (D) * 1052 (H) mm |
Mass | 320 Kg |
Power supply | AC 200 V ± 10 % |
Power consumption | 1.5 KVA, Maximun |